FEEDBACK
PLEASE WAIT ...
CLOSE
CLOSE

Manufacturer

Country

Registration Date 9 Nov 2016
Share

TiGRT MLC

(0)

Others Electronics and Appliances

Multileaf Collimator

Applications

Conformal Radiation Treatment Intensity Modulated Radiation Treatment

Properties

MLCs are easily integrated with existing linear accelerators and can be used for the majority of patient treatments. The multiple 1.6mm physical thickness leaves at isocenter give the MLC the ability to accurately and precisely treat small fields using either CRT or IMRT protocols. The MLC provides less than 2% inter-leaf leakage. RVS provides automatic initialization at program start up and a self-calibration protocol. Remote diagnosis, redundancy design and maintenance functions offer a user-friendly system with low operating costs and enhanced reliability.

Manufacturer's Description

The TiGRT MLC is designed using advanced DSP and FPGA electronic design technologies and incorporates our user friendly TiGRT TPS software to provide the user with the ability to implement both Conformal Radiation Treatment (CRT), Intensity Modulated Radiation Treatment (IMRT) as well as Sliding ARC protocols, and works with TiGRT IVS to perform SRS and 4D tracking functions. All functions are available in one fully integrated treatment package!
Model: MLC M / HLeaf Number: 54 / 102Mini Leaf Thickness: 1.6mmMax Field Size: 300mm×150mm / 300mm×300mm depend on Linac type Max Over Travel: 50mm Max Leaf Speed: 50mm/s Weight: 39kg / 65kg Interleaf Leakage: <2% Position Error: <0.5mm Penumbra: <4mm Adapter Type: All major linac, C60, Proton Machines; Application:Conformal SRS/SRT/SBRT Step & Shoot IMRT Dynamic Sliding IMRT Sliding ARC IMRT VMAT, 4D tracking